Ad-weighing SV-100 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Ad-weighing SV-100. A&D Weighing SV-100 User Manual [en] Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Vibro Viscometer

SV-10 SV-100 Vibro Viscometer INSTRUCTION MANUAL 1WMPD4000646E

Strany 2

83. DISPLAY AND KEYS 3-1 Display Name Description Standby mode Displays [- - - - -]. Measurement mode Displays the viscosity value in

Strany 3 - CONTENTS

93-2 Keys Key Description ON:OFF Power Turns the power on and off. When the power is turned on, the viscometer enters the standby mode ( [- -

Strany 4

103-3 Displaying the Viscosity Values The viscosity values are displayed as below, depending on the unit selected and the viscosity range. The corre

Strany 5 - 1. INTRODUCTION

11 3-3-2 SV-100 Use the MODE key to switch between Pa⋅s (Pascal second) and P (Poise). The unit selected at the factory before shipment is Pa⋅s. U

Strany 6

124. PRECAUTIONS To get the optimum performance from the viscometer and acquire accurate measurement data, note the following: 4-1 General Precautio

Strany 7 - 1-2 Features

134-2 During Use  To level the surface of the sample; adjust the leveling feet so that the center of the narrow part of the right and left sensor pl

Strany 8 - 2. UNPACKING THE VISCOMETER

144-3 After Use  Remove any residual sample material from the sensor plates, temperature sensor and protector using alcohol. Using the sensor plates

Strany 9

154-4 Measuring the Absolute Value of Viscosity The SV Series Sine-wave Vibro Viscometer, as a measuring principle, detects the product of viscosity

Strany 10 - 3. DISPLAY AND KEYS

164-4-2 At Calibration When calibrating, enter the product of the absolute viscosity value and the density of the standard viscosity fluid used for c

Strany 11 - 3-2 Keys

175. MEASUREMENT 5-1 Preparing the Sample 1 Pour the sample into the cup until its surface reaches between the level gauges. The level gauges indic

Strany 12 - 3-3-1 SV-10

© 2008 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any l

Strany 13 - 3-3-2 SV-100

18NO Note Use the protector in the position as shown on the left below. If the protector is not used with the SV-10, a measurement error may occur,

Strany 14 - 4. PRECAUTIONS

195-2 Basic Measurement Procedure 1 With the power turned off, press the ON:OFF key. When the display is in the standby mode, press the START ke

Strany 15 - 4-2 During Use

205-3 Changing Units The units of viscosity available are: mPa⋅s (millipascal second), Pa⋅s (Pascal second), cP (Centipoise), P (Poise). The cor

Strany 16 - 4-3 After Use

216. VISCOSITY CALIBRATION  Viscosity calibration is to correct the viscosity value.  Two calibration methods are available: one-point calibration

Strany 17 - 4-4-1 At Measurement

226-2 Calibration Procedure Note As to the correction value used for one-point calibration and two-point calibration, enter the product of the absol

Strany 18 - 4-4-2 At Calibration

236-2-1 One-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Selec

Strany 19 - 5. MEASUREMENT

246-2-2 Two-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Sele

Strany 20

25 9 When the measurement of the first point has completed, clean the sensor plates, temperature sensor and protector and prepare the second standar

Strany 21

266-2-3 Simplified Calibration Using Purified Water (SV-10 only) 1 Place the purified water in the sample cup. 2 Press the START key to measure

Strany 22 - 5-3 Changing Units

27 Reference data: Theoretical viscosity value (Viscosity × Density) of the purified water at various temperatures Temperature (°C) Viscosity × D

Strany 23 - 6. VISCOSITY CALIBRATION

1CONTENTS 1. INTRODUCTION ...

Strany 24 - 6-2 Calibration Procedure

287. FUNCTION SETTING The viscometer, by selecting functions to be used in the function setting, can specify the performance appropriate to the usage

Strany 25 - 6-2-1 One-point Calibration

29 Example of the function setting procedure Using the SV-10, to change the setting of "Unit upon power-on (Unit)" to the viscosity: cP (Ce

Strany 26 - 6-2-2 Two-point Calibration

307-2 Details of the Function Items Function item Parameter Description Cladj Date/Time Sets the order of the date (YMD,MDY,DMY) and the date/ti

Strany 27

317-3 Description of Items Date/Time (Cladj)  The upper two digits of the year are not displayed. For example, the year 2003 is displayed as "

Strany 28 - of the

32 Changing the date The date is changed in the selected displaying order. The following is an example when the order of " y " (Year), &quo

Strany 29

33 Changing the time 12 The current time is displayed. Press the MODE key to display the current date. 13 Press the PRINT key to select the setti

Strany 30 - 7. FUNCTION SETTING

34 Condition (Cond) The stability of the viscosity measurement results can be adjusted, taking ambient conditions such as vibration into consideratio

Strany 31

35 Decimal Point (pnt) Parameter Settings Description 0 • Dot "." 1 Comma "," The decimal point format for the displayed measu

Strany 32

36 Data Output Format (type) The output format appropriate for the device connected to RS-232C can be selected. Parameter Settings Description 0 A

Strany 33 - 7-3 Description of Items

37 Date/time Output (5-td) Parameter Settings Description 0 No output 1 • Output With D.P. format or CSV format selected, whether or not to add the

Strany 34

2 8. CONNECTION TO A PERSONAL COMPUTER...49 9. CONNECTION TO A PRINTER ...

Strany 35

38 Setting the ID number 1 In the standby mode, press and hold the MODE key to enter the function setting mode. " Cladj " appears. 2 Pr

Strany 36

39 Initialization (Clr) Restores the following data to the default setting.  Function setting  Calibration data After initialization, check the vis

Strany 37

407-4 Data Output Format Examples 7-4-1 A&D Standard Format Used with the printer MODE 1 or MODE 2 when the optional compact printer, AD-8121B

Strany 38

41 SV-100 output format example Viscosity unit Display Output format Remarks L Pa・s OL,-99999999Pas Below measuring range error 1.00 Pa ・s

Strany 39 - Display

427-4-2 D.P. Format Used with the printer MODE 3 when the optional compact printer, AD-8121B is connected. With "prt 0" or "prt 1&qu

Strany 40

43 Printing format example (4) Function setting (√=Output Blank=No output) 5-at 1 Measurement elapsed time √ 5-td 0 Date/time Remarks Device ID i

Strany 41 - Initialization (Clr)

44 Output format example (1) With ID number, date and time added Function setting (√=Output Blank=No output) 5-td 1 Date/time √ 5-ed 3 Device ID n

Strany 42

45 SV-100 output format example Viscosity / Temper-ature Display Output format example Remarks L Pa・s LAB-12,2003/03/19,12:34:56,+025.67,C,+0

Strany 43

46 Output format example (2) With date and time added Function setting (√=Output Blank=No output) 5-td 1 Date/time √ 5-ed 0 Device ID number SV-

Strany 44 - 7-4-2 D.P. Format

477-4-4 RsVisco Format Used with the graphing program RsVisco. Measurement data are output in comma separated format. When a comma is selected as the

Strany 45 - 7-4-3 CSV Format

31. INTRODUCTION This manual describes how the SV series viscometer works and how to get the most out of it in terms of performance. Read this manu

Strany 46

48 SV-100 output format example Viscosity / Temper-ature Display Output format example Remarks L Pa・s +0000.000, Pa s,+025.67,C Zeroes are o

Strany 47

498. CONNECTION TO A PERSONAL COMPUTER Using the standard RS-232C serial interface and the RS-232C cable provided, the viscometer can be connected t

Strany 48 - ,,,+025.67,C,+00001.23, Pa s

509. CONNECTION TO A PRINTER  The viscometer can be connected to the optional compact printer, AD-8121B using the standard RS-232C interface and th

Strany 49 - 7-4-4 RsVisco Format

5110. SERIAL INTERFACE RS-232C Serial Interface Transmission system EIA RS-232C Transmission form Asynchronous, bi-directional, half duplex Data fo

Strany 50

5211. COMMAND LIST The viscometer can be controlled by the following commands from the computer. Add a terminator ( 0Dh、0Ah ) to each command. Comm

Strany 51

5312. TROUBLESHOOTING The viscometer is a precision instrument. When the measuring environment or the measurement method is inadequate, correct value

Strany 52 - 9. CONNECTION TO A PRINTER

54How to remove and attach the surface locator plate Removing Loosen the screw and remove the surface locator plate from the temperature sensor. A

Strany 53 - 10. SERIAL INTERFACE

55When measurement values are not correct:  Has the sample surface been adjusted to the center of the narrow part of the sensor plates?  Adjust th

Strany 54 - 11. COMMAND LIST

56 When more precise measurement is required:  When the viscometer is installed for the first time or is moved to another location, plug in the AC a

Strany 55 - 12. TROUBLESHOOTING

5713. ERROR DISPLAY Error display Description Above measuring range error The viscosity value exceeds the upper limit of the viscosity measuring r

Strany 57

5814. SPECIFICATIONS Measurement method SV-10 SV-100 Measurement method Sine-wave Vibro Viscometer using the Tuning-fork Vibration method Vibration

Strany 58

59 *1 When a sample cup of 45 mL is used. *2 Repetitive measurement with the sensor plates remaining in the sample *3 The value after calibratio

Strany 59 - 13. ERROR DISPLAY

6015. OPTIONAL ACCESSORIES List of Optional Accessories (sold separately) Name Number Standard viscosity fluid (JS2.5) *1 AX-SV-31-2.5 Standard

Strany 60 - 14. SPECIFICATIONS

61AX-SV-35 Glass Sample Cup (Approx. 13 mL) • Used when organic solvents will be measured. • The AX-SV-37 consists of : Glass sample cup (Capacity:

Strany 61

62AX-SV-37 Water Jacket Assembly • Used, in combination with a commercially available constant temperature bath for heating medium circulation, to m

Strany 62 - 15. OPTIONAL ACCESSORIES

63 AX-SV-38 Glass Storage Container • Used to store sample fluids. • The viscosity of the sample stored in the container can be measured as it

Strany 63 - AX-SV-36 Positioning Stopper

6416. EXTERNAL DIMENSIONS Whole View Unit: mm Detailed View of the Sensor Unit Unit: mm

Strany 64

51-2 Features  High accuracy The Sine-wave Vibro Viscometer achieves a high measurement accuracy of 1%*1 (repeatability) over the full range. *1 R

Strany 65 - Glass storage container

62. UNPACKING THE VISCOMETER 2-1 Unpacking  The viscometer is a precision instrument. Unpack the viscometer carefully. Keep the packing material t

Strany 66 - 16. EXTERNAL DIMENSIONS

72-2 Installing the Viscometer Install the viscometer as follows: 1 Connect the display unit to the main unit using the connection cable.

Příbuzné modely SV-10

Komentáře k této Příručce

Žádné komentáře